Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

On the whole, the system appears to be is capable of high sensitivity imaging in Stokes I (see commissioning reports below). We note that polarisation leakage is significant in the upper half of the band and at this stage polarimetric observations in the S3 and S4 sub-bands are not advised - see discussion below.

...

Info

Switching between sub-bands requires 'rebuilding' of the sub-array. Including time for delay calibration and subsequent operator quality assurance checks, this process takes about 15 - 20 minutes. Note that if wider frequency coverage is required in the same time slot, but sensitivity is not as vital, multiple sub-arrays with a subset of antennas can be run simultaneously in different bands.

Sensitivity

The mean SEFD SEFDs per antenna is are given in Figure Figures 1 and 2 below. Note that these are preliminary results, yet to be refined by gain curves.

...